Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607

On 2018-04-25 the FDA classified a Class II recall of Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607 by Cook, citing pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1460-2018
ClassificationClass II
Statusterminated
Initiated2018-03-22
Published2018-04-25
Last updated2026-09-13 11:54 UTC
CompanyCook
Distributionn/a

Product

Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.

Reason

Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

Identifiers

code_infoAll lots