RecallRadar

FDA Device recall Z-1457-2015

ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894

On 2015-04-29 the FDA classified a Class II recall of ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894 by Siemens Healthcare Diagnostics, citing siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier result….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1457-2015
ClassificationClass II
Statusterminated
Initiated2015-03-23
Published2015-04-29
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.

Reason

Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.

Identifiers

code_infoCatalog Number: B01-4840-01; Material Number (SMN)10311896, Lot Numbers: 318706, 332463, expiration dates 2015-06-30, 2015-08-31

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1457-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.