RecallRadar

FDA Device recall Z-1457-2013

Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable S…

On 2013-06-12 the FDA classified a Class I recall of Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable S… by Symbios Medical Products, citing symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured w….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1457-2013
ClassificationClass I
Statusterminated
Initiated2013-05-10
Published2013-06-12
Last updated2026-09-13 11:53 UTC
CompanySymbios Medical Products
DistributionUS

Product

Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.

Reason

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Identifiers

code_infoPump Part 510449, Lot 11-100271 - Lot 130100137 found in: GoBlock Kit 510472, Lot 11-100385 to 13-100354

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1457-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.