RecallRadar

FDA Device recall Z-1454-2013

Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated cathete…

On 2013-06-12 the FDA classified a Class I recall of Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated cathete… by Symbios Medical Products, citing symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured w….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1454-2013
ClassificationClass I
Statusterminated
Initiated2013-05-10
Published2013-06-12
Last updated2026-09-13 11:53 UTC
CompanySymbios Medical Products
DistributionUS

Product

Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated catheter & BIOPATCH; Go Pump Kit 510349-BP, Epidural Catheter, BioPatch; Go Pump Kit 510642 Disposable Single Infusion Pump, Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. The medication is delivered via tubing to the surgical site at a very slow rate

Reason

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Identifiers

code_info
Pump Part 510180, Lot 11-100453 to 12-101003 found in: Go Pump Kit 510201-BP. Lot 12-100108 to 12-101538 Go Pump Ki…Pump Part 510180, Lot 11-100453 to 12-101003 found in: Go Pump Kit 510201-BP. Lot 12-100108 to 12-101538 Go Pump Kit 510349-BP. Lot 12-100935 Go Pump Kit 510642, Lot 12-100616 to 13-100590

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1454-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.