FDA Device recall Z-1453-2022
Description/Part: Simplify Disc Size 1, Ht 5/SM-5
- FDA Device
- Class II
- ongoing
- Published 2022-08-03
On 2022-08-03 the FDA classified a Class II recall of Description/Part: Simplify Disc Size 1, Ht 5/SM-5 by Nuvasive, citing cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1453-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-06-20 |
| Published | 2022-08-03 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Nuvasive |
| Distribution | CA, DE, FL, IN, MI, MO, OR, TX, VA, WA |
Product
Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5/MD-5L; Simplify Disc Size 2, Ht 6/MD-6; Simplify Disc Size 2, Ht 6, 5/MD-6L; Simplify Disc Size 3, Ht 5/LG-5; Simplify Disc Size 3, Ht 5, 5/LG-5L; Simplify Disc Size 3, Ht 6/LG-6; Simplify Disc Size 3, Ht 6, 5/LG-6L
Reason
Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.
Identifiers
| code_info | Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132,…Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132, W81133, W81633, W81625, W81611, W81828, W81971, W82544, W82543, W81034, W81975, W81580, W81612, W81621, W81131, W81613, W80909, W81623, W81634, W79466, W81033, W81134, W81600, W79259; MD-5/843285100061/W81136, W81135, W80704, W80700; MD-5L/843285100078/W80707, W80706; MD-6/843285100085/W80911, W81642, W81643, W82545, W82546; MD-6L/843285100092/W81822; LG-5/843285100108/W81138, W81141, W79465, W81994, W81142, W81626, W81143, W79257; LG-5L/843285100115/W79258, W80146, W80915, W82548; LG-6/843285100122/W82547; LG-6L/843285100139/W81827 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1453-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.