FDA Device recall Z-1451-2015
Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Produc…
- FDA Device
- Class II
- terminated
- Published 2015-04-22
On 2015-04-22 the FDA classified a Class II recall of Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Produc… by Baxter Healthcare, citing loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the near….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1451-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-02 |
| Published | 2015-04-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Baxter Healthcare |
| Distribution | CA, CT, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NH, NY, PR, SD, TX, UT, WV |
Product
Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Product Code 35700BAX2 The SIGMA Spectrum Infusion Pump with Master Drug Library (MDL) is intended to be used for the controlled administration of fluids.
Reason
Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.
Identifiers
| code_info | Software version 8.0, Product Code 35723V080 |
|---|