FDA Device recall Z-1440-2013
Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 3…
- FDA Device
- Class II
- terminated
- Published 2013-06-12
On 2013-06-12 the FDA classified a Class II recall of Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 3… by Carl Zeiss Meditec, citing carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1440-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-04-18 |
| Published | 2013-06-12 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Carl Zeiss Meditec |
| Distribution | US |
Product
Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C), Product Usage: A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.
Reason
Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.
Identifiers
| code_info | Serial Numbers: 6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813 |
|---|