RecallRadar

FDA Device recall Z-1439-2017

CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15

On 2017-03-22 the FDA classified a Class II recall of CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15 by Smiths Medical Asd, citing 20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1439-2017
ClassificationClass II
Statusterminated
Initiated2016-10-31
Published2017-03-22
Last updated2026-09-13 11:54 UTC
CompanySmiths Medical Asd
Distributionn/a

Product

CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,

Reason

20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.

Identifiers

code_info
1071155, 1076826, 1081258, 1081260, 1081275, 1082475, 1082511, 1082512, 1090347, 1090348, 1090349, 1091767, 1091768, 10…1071155, 1076826, 1081258, 1081260, 1081275, 1082475, 1082511, 1082512, 1090347, 1090348, 1090349, 1091767, 1091768, 1091769, 1091770, 1091771, 1092395, 1100186, 1100188, 1100189.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1439-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.