FDA Device recall Z-1435-2013

Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring…

On 2013-06-05 the FDA classified a Class II recall of Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring… by Heart Sync, citing on 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage d….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1435-2013
ClassificationClass II
Statusterminated
Initiated2013-04-26
Published2013-06-05
Last updated2026-09-13 11:53 UTC
CompanyHeart Sync
DistributionAL, GA, MA, MI, MN, NC, ND, OH, PA, SD, TX, VA, VT

Product

Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Reason

On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

Identifiers

code_infoPart Number: D304 and lot numbers:Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01