FDA Device recall Z-1431-2013
8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200
- FDA Device
- Class II
- terminated
- Published 2013-06-05
On 2013-06-05 the FDA classified a Class II recall of 8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200 by Intuitive Surgical, citing reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery fun….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1431-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-05-08 |
| Published | 2013-06-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
Reason
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Identifiers
| code_info | Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M1…Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M10121115 M10121123 M10121130 M10121206 M10121210 M10130124 M10130131 M10130205 M10130304 M10130319 M10130321 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1431-2013%22