RecallRadar

FDA Device recall Z-1428-2017

Apollo

On 2017-03-15 the FDA classified a Class II recall of Apollo by Villa Sistemi Medicali S P A, citing patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1428-2017
ClassificationClass II
Statusterminated
Initiated2016-12-27
Published2017-03-15
Last updated2026-09-13 11:54 UTC
CompanyVilla Sistemi Medicali S P A
DistributionUS

Product

Apollo; Models: 9784000131 9784000231 9784020131 9784020231 9784100231 9784120131 9784120231 9784130131 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.

Reason

Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Identifiers

code_infoSerial Numbers: 06100127 06110138 07010149 07020158 07020163 07020165 07030171 07040206 09010566 09090605 10100752 11100873 15041446

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1428-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.