FDA Device recall Z-1427-2017

Juno DRF

On 2017-03-15 the FDA classified a Class II recall of Juno DRF by Villa Sistemi Medicali S P A, citing patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1427-2017
ClassificationClass II
Statusterminated
Initiated2016-12-27
Published2017-03-15
Last updated2026-09-13 11:54 UTC
CompanyVilla Sistemi Medicali S P A
DistributionUS

Product

Juno DRF; Model: 709020 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures

Reason

Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Identifiers

code_info
Serial Numbers: 10110770 11030805 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 120…Serial Numbers: 10110770 11030805 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487 15121505 16011512 16021514 16021517 16021518 16041533 16121595

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1427-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.