RecallRadar

FDA Device recall Z-1426-2024

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of huma…

On 2024-04-10 the FDA classified a Class II recall of Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of huma… by Life Technologies, citing torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1426-2024
ClassificationClass II
Statusongoing
Initiated2024-02-07
Published2024-04-10
Last updated2026-09-13 11:57 UTC
CompanyLife Technologies
DistributionUS

Product

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A33178/ A33178 (5)Torrent Suite Dx Software 5.0- A29166/ A29166

Reason

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings, configurations, software, or data on the instrument

Identifiers

code_infoUDI: 10190302017848;10190302016810 Torrent Suite Dx Software versions 5.14 and earlier

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1426-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.