RecallRadar

FDA Device recall Z-1425-2022

Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology s…

On 2022-07-27 the FDA classified a Class II recall of Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology s… by Siemens Medical Solutions Usa, citing sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1425-2022
ClassificationClass II
Statusongoing
Initiated2022-06-15
Published2022-07-27
Last updated2026-09-13 11:56 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model:10764561

Reason

Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.

Identifiers

code_infoUDI-DI: 04056869010137 Serial Numbers: 61331, 60015