FDA Device recall Z-1421-2024
Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty
- FDA Device
- Class II
- ongoing
- Published 2024-04-10
On 2024-04-10 the FDA classified a Class II recall of Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty by Exactech, citing the packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1421-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-03-06 |
| Published | 2024-04-10 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Exactech |
| Distribution | US |
Product
Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty
Reason
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
Identifiers
| code_info | a) 320-40-10, UDI/DI 10885862535054, Serial Numbers: 5937242, 5910691, 5937233, 5936848, 6062040, 5936849, 6197662, 61…a) 320-40-10, UDI/DI 10885862535054, Serial Numbers: 5937242, 5910691, 5937233, 5936848, 6062040, 5936849, 6197662, 6197672, 6197682, 6197680, 6197683, 6197674, 6413423, 6062107, 6200990, 6046848, 6197006, 6062047, 5910110, 6413379, 6198008, 6197004, 5936902, 6484655, 5937219, 6197676, 5936834, 6533569, 6197660, 5936882, 6533590, 5910120, 6046849, 6046836, 6062039, 6046843, 6046847, 6413376, 6533602, 6533603, 6533605, 6533593, 6533595, 6533596, 6197026, 6484652, 6484654, 6484665, 6484672, 6198014, 6646211, 5910690, 5936906, 6198016, 6046852, 5937248, 5937208, 6533421, 5936873, 6533292, 6200998, 5936825, 6200968, 6533303, 5910109, 6197681, 5937249, 5937216, 6413403, 6198001, 6197003, 6533302, 6200966, 5937235, 6062099, 6062068, 6197025, 6062095, 6646201, 6673947, 6413375, 6062034, 5937245, 6198002, 6062042, 6413412, 5936866, 6062096, 5910123, 6062051, 6062036, 5936910, 6200985, 5936878, 6200982, 6533422, 6413386, 6197667, 5936869, 6197005, 6200976, 6413408, 6197663, 5937234, 6200973, 6 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1421-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.