RecallRadar

FDA Device recall Z-1421-2024

Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty

On 2024-04-10 the FDA classified a Class II recall of Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty by Exactech, citing the packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1421-2024
ClassificationClass II
Statusongoing
Initiated2024-03-06
Published2024-04-10
Last updated2026-09-13 11:57 UTC
CompanyExactech
DistributionUS

Product

Exactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty

Reason

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Identifiers

code_info
a) 320-40-10, UDI/DI 10885862535054, Serial Numbers: 5937242, 5910691, 5937233, 5936848, 6062040, 5936849, 6197662, 61…a) 320-40-10, UDI/DI 10885862535054, Serial Numbers: 5937242, 5910691, 5937233, 5936848, 6062040, 5936849, 6197662, 6197672, 6197682, 6197680, 6197683, 6197674, 6413423, 6062107, 6200990, 6046848, 6197006, 6062047, 5910110, 6413379, 6198008, 6197004, 5936902, 6484655, 5937219, 6197676, 5936834, 6533569, 6197660, 5936882, 6533590, 5910120, 6046849, 6046836, 6062039, 6046843, 6046847, 6413376, 6533602, 6533603, 6533605, 6533593, 6533595, 6533596, 6197026, 6484652, 6484654, 6484665, 6484672, 6198014, 6646211, 5910690, 5936906, 6198016, 6046852, 5937248, 5937208, 6533421, 5936873, 6533292, 6200998, 5936825, 6200968, 6533303, 5910109, 6197681, 5937249, 5937216, 6413403, 6198001, 6197003, 6533302, 6200966, 5937235, 6062099, 6062068, 6197025, 6062095, 6646201, 6673947, 6413375, 6062034, 5937245, 6198002, 6062042, 6413412, 5936866, 6062096, 5910123, 6062051, 6062036, 5936910, 6200985, 5936878, 6200982, 6533422, 6413386, 6197667, 5936869, 6197005, 6200976, 6413408, 6197663, 5937234, 6200973, 6

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1421-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.