FDA Device recall Z-1420-2022
Ambu VivaSight 2 DLT
- FDA Device
- Class II
- ongoing
- Published 2022-07-27
On 2022-07-27 the FDA classified a Class II recall of Ambu VivaSight 2 DLT by Ambu, citing firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1420-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-06-03 |
| Published | 2022-07-27 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ambu |
| Distribution | US |
Product
Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000, 412391000, 412411000
Reason
Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.
Identifiers
| code_info | Catalog No: 412351000 UDI-DI: 5707480145706 1 each 5707480145713 5 pc/pack Catalog No. 412371000 UDI-DI: 570…Catalog No: 412351000 UDI-DI: 5707480145706 1 each 5707480145713 5 pc/pack Catalog No. 412371000 UDI-DI: 5707480145737 1 each 5707480145744 5 pc/pack Catalog No. 412391000 UDI-DI: 5707480145768 1 each 5707480145775 5 pc/pack Catalog No. 412411000 UDI-DI: 5707480145799 1 each 5707480145805 5 pc/pack All lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1420-2022%22