RecallRadar

FDA Device recall Z-1420-2018

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

On 2018-04-25 the FDA classified a Class II recall of Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B by Autochair, citing recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1420-2018
ClassificationClass II
Statusongoing
Initiated2018-03-07
Published2018-04-25
Last updated2026-09-13 11:54 UTC
CompanyAutochair
DistributionPA

Product

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Reason

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Identifiers

code_infoAll Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1420-2018%22