FDA Device recall Z-1420-2018
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
- FDA Device
- Class II
- ongoing
- Published 2018-04-25
On 2018-04-25 the FDA classified a Class II recall of Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B by Autochair, citing recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1420-2018 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2018-03-07 |
| Published | 2018-04-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Autochair |
| Distribution | PA |
Product
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
Reason
Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
Identifiers
| code_info | All Lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1420-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.