FDA Device recall Z-1419-2023
Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs
- FDA Device
- Class II
- ongoing
- Published 2023-04-26
On 2023-04-26 the FDA classified a Class II recall of Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs by Beckman Coulter, citing the reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1419-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-03-15 |
| Published | 2023-04-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.
Reason
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.
Identifiers
| code_info | Lot #233968, exp. 2023-06-30, UDI (01)15099590201661(17)230630(11)220630(10)233968. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1419-2023%22