RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0

On 2013-06-05 the FDA classified a Class II recall of RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0 by Raysearch Laboratories Ab, citing under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1419-2013
ClassificationClass II
Statusterminated
Initiated2013-02-15
Published2013-06-05
Last updated2026-09-13 11:53 UTC
CompanyRaysearch Laboratories Ab
DistributionCT, FL, MA, MO, NY, OH

Product

RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason

Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.

Identifiers

code_info
Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bri…Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bring up the About box in the RayStation application and check the build number reported0.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1419-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.