FDA Device recall Z-1418-2026
Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I
- FDA Device
- Class II
- ongoing
- Published 2026-02-25
On 2026-02-25 the FDA classified a Class II recall of Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I by Medline Industries, citing medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packa….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1418-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-01-07 |
| Published | 2026-02-25 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807K; 5) MINOR VAGINAL #76-RF, Model Number: DYNJ27434R; 6) VAG HYST PACK, Model Number: DYNJ31344J; 7) VAG HYST PACK, Model Number: DYNJ31344K; 8) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393L; 9) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393M; 10) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393N; 11) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393O; 12) LAVH PACK, Model Number: DYNJ36712F; 13) D&C MINOR LITHOTOMY PACK, Model Number: DYNJ45090I; 14) D C HYSTEROSCOPY PACK-LF, Model Number: DYNJ45801C; 15) VAG HYST PACK (VHGSG)642-LF, Model Number: DYNJ47694I; 16) HYSTEROSCOPY D AND C PACK, Model Number: DYNJ51741B; 17) LAVH - N #668629 -N, Model Number: DYNJ58121C; 18) PK, GYN-MINOR-LITHO, Model Number: DYNJ59037B; 19) HYSTEROSCOPY PACK, Model Number: DYNJ59444G; 20) D AND C PACK, Model Number: DYNJ61282B; 21) VAGINAL HYSTERECTOMY, Model Number: DYNJ67161; 22) PERI GYN, Model Number: DYNJ67708A; 23) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713B; 24) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713C; 25) ANORECTAL PACK, Model Number: DYNJ83560; 26) D&C CDS, Model Number: DYNJ902560I; 27) HYSTEROSCOPY, Model Number: DYNJ903327K; 28) ROBOTIC HYST, Model Number: DYNJ903798G; 29) TLH KIT, Model Number: DYNJ908149; 30) LAVH TOTE, Model Number: DYNJ908340; 31) LAVH TOTE, Model Number: DYNJ908340A; 32) LAVH TOTE, Model Number: DYNJ908340B; 33) LAVH TOTE, Model Number: DYNJ908340D; 34) BRECKENRIDGE LAVH, Model Number: DYNJ910537; 35) VAG HYST PACK, Model Number: DYNJT3348; 36) LAVH PART B, Model Number: DYNJV0312G
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Identifiers
| code_info | 1) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 24CBQ563; 2) CDS984086I, UDI-DI: 1019532…1) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 24CBQ563; 2) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 24ABO363; 3) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 23IBV204; 4) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 23HBY878; 5) CDS984086I, UDI-DI: 10195327428006(each), 40195327428007(case), Lot Number: 23GBE705; 6) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 23HBI773; 7) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 23FBD256; 8) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 23FBC116; 9) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 23CBS135; 10) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 23CBL440; 11) DYKMBNDL91A, UDI-DI: 10195327246891(each), 40195327246892(case), Lot Number: 22KBH451; 12) DYKMBNDL9 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.