RecallRadar

FDA Device recall Z-1417-2021

MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product…

On 2021-04-21 the FDA classified a Class II recall of MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product… by Vectornate Usa, citing discoloration and turbidity were present in the specified lots of STMs because of Longer Monsoon season at the South Korean Manufacturing Facility.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1417-2021
ClassificationClass II
Statusterminated
Initiated2021-02-16
Published2021-04-21
Last updated2026-09-13 11:56 UTC
CompanyVectornate Usa
DistributionUS

Product

MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product Usage: intended for the collection and transport of clinical specimens containing viruses, from the collection site to the testing laboratory. I

Reason

Discoloration and turbidity were present in the specified lots of STMs because of Longer Monsoon season at the South Korean Manufacturing Facility.

Identifiers

code_infoSerial #: STM-UNV-0001-Q121, Lot #: JAH161, KAG076

Status history