FDA Device recall Z-1415-2023
BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform
- FDA Device
- Class II
- ongoing
- Published 2023-04-26
On 2023-04-26 the FDA classified a Class II recall of BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform by Becton Dickinson And, citing upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occ….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1415-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-03-22 |
| Published | 2023-04-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro diagnostic medical device intended to automate specimen processing according to user-defined procedures and protocols. Catalog (Ref) No. : 447213
Reason
Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.
Identifiers
| code_info | UDI: 038290LSUVLPKFPV Serial Numbers: INO-000156, INO-000249, INO-000248, INO-000252, INO-000254, 20130276, 19033ST002 |
|---|