FDA Device recall Z-1415-2021
The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose
- FDA Device
- Class II
- terminated
- Published 2021-04-21
On 2021-04-21 the FDA classified a Class II recall of The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose by Eos Imaging, citing inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1415-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-12-22 |
| Published | 2021-04-21 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Eos Imaging |
| Distribution | US |
Product
The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. ¿ The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). ¿ The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.
Reason
Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
Identifiers
| code_info | Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.19…Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.1909.1006, 8.2005.1014, 8.1905.1004, 8.1911.1007, 8.1912.1009, and 8.2006.1016 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1415-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.