RecallRadar

FDA Device recall Z-1414-2017

EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Produc…

On 2017-03-15 the FDA classified a Class II recall of EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Produc… by Becton Dickinson And, citing the information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1414-2017
ClassificationClass II
Statusterminated
Initiated2014-08-25
Published2017-03-15
Last updated2026-09-13 11:54 UTC
CompanyBecton Dickinson And
Distributionn/a

Product

EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.

Reason

The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.

Identifiers

code_infoLot Numbers Version 6.20A Cat. Number 442953

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1414-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.