RecallRadar

FDA Device recall Z-1413-2017

RSP Impaction Fixture

On 2017-03-15 the FDA classified a Class II recall of RSP Impaction Fixture by Encore Medical, citing during the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titani….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1413-2017
ClassificationClass II
Statusterminated
Initiated2017-02-09
Published2017-03-15
Last updated2026-09-13 11:54 UTC
CompanyEncore Medical
Distributionn/a

Product

RSP Impaction Fixture

Reason

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Identifiers

code_info109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1413-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.