FDA Device recall Z-1412-2023
INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026
- FDA Device
- Class II
- ongoing
- Published 2023-04-26
On 2023-04-26 the FDA classified a Class II recall of INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026 by Medtronic Sofamor Danek Usa, citing incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1412-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-02-16 |
| Published | 2023-04-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Medtronic Sofamor Danek Usa |
| Distribution | US |
Product
INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.
Reason
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Identifiers
| code_info | GTIN 00643169554696, Lot/Batch numbers: H5806345 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Medtronic Sofamor Danek Usa
- Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone ma…
FDA Device · Class II · 2026-10-21 · Medtronic Sofamor Danek Usa - Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone ma…
FDA Device · Class II · 2026-10-21 · Medtronic Sofamor Danek Usa - Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM…
FDA Device · Class II · 2025-12-03 · Medtronic Sofamor Danek Usa - Grafton DMB Matrix, Model Numbers: a) T42200, b) T42210
FDA Device · Class II · 2023-11-01 · Medtronic Sofamor Danek Usa - Grafton DBM Putty, Model Numbers: a) T43102, b) T43102INT, c) T43103INT, d) T43105, e) T43105INT, f) T43110, g) T43110INT
FDA Device · Class II · 2023-11-01 · Medtronic Sofamor Danek Usa
All Medtronic Sofamor Danek Usa recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1412-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.