RecallRadar

FDA Device recall Z-1410-2021

Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catal…

On 2021-04-21 the FDA classified a Class II recall of Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catal… by K2m, citing interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserte….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1410-2021
ClassificationClass II
Statusterminated
Initiated2021-03-16
Published2021-04-21
Last updated2026-09-13 11:56 UTC
CompanyK2m
DistributionPA

Product

Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033

Reason

Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies

Identifiers

code_infoLot Number: MNAJ GTIN: 10888857082571

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1410-2021%22