FDA Device recall Z-1408-2020

McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-00…

On 2020-03-11 the FDA classified a Class II recall of McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-00… by Medtronic, citing blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video scre….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1408-2020
ClassificationClass II
Statusongoing
Initiated2020-01-28
Published2020-03-11
Last updated2026-09-13 11:55 UTC
CompanyMedtronic
DistributionUS

Product

McGRATH X3 Disposable Laryngoscope Blades-used with the McGRATH MAC Video Laryngoscope to aid the intubation of the trachea Code: X3-002-000 (Individual Unit); X3-003-000 (Carton of 10 Units) - Product Usage: is a tool used to aid the intubation of the trachea. As a rigid laryngoscope it holds and shapes the anatomy allowing a clear view of the larynx and entrance to the trachea.

Reason

Blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video screen

Identifiers

code_infoUDI: 15060272980082 Lot/Serial #: M1811191 M1811261 M1812071 M1812271 M1901081 M1901161 M1902251 M1903131 M1904031 M1904241 M1906031 M1906121

Status history