FDA Device recall Z-1402-2022

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part num…

On 2022-07-27 the FDA classified a Class II recall of Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part num… by Stryker Instruments Div, citing some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Cani….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1402-2022
ClassificationClass II
Statusterminated
Initiated2020-01-22
Published2022-07-27
Last updated2026-09-13 11:56 UTC
CompanyStryker Instruments Div
DistributionUS

Product

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

Reason

Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.

Identifiers

code_infoUDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24