FDA Device recall Z-1400-2025
VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1
- FDA Device
- Class II
- ongoing
- Published 2025-03-26
On 2025-03-26 the FDA classified a Class II recall of VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1 by Merge Healthcare, citing when following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clin….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1400-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-02-19 |
| Published | 2025-03-26 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Merge Healthcare |
| Distribution | n/a |
Product
VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Reason
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Identifiers
| code_info | Lot Code: Software Version Numbers and UDIs 11.0.2: UDI: (01)00842000100416(10)11.0.2.1523(11)180809; 11.0.3: UDI:…Lot Code: Software Version Numbers and UDIs 11.0.2: UDI: (01)00842000100416(10)11.0.2.1523(11)180809; 11.0.3: UDI: (01)00842000100416(10)11.0.3.1601(11)181019; 11.0.4: UDI: (01)00842000100416(10)11.0.4(11)210207; 11.1: UDI: (01)00842000100751(10)11.1.0.1054(11)181211; 11.1.1: UDI: (01)00842000100751(10)11.1.1(11)190724 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1400-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.