RecallRadar

FDA Device recall Z-1395-2023

GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

On 2023-04-26 the FDA classified a Class II recall of GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System by Gramercy Extremity Orthopedics, citing instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1395-2023
ClassificationClass II
Statusongoing
Initiated2023-04-04
Published2023-04-26
Last updated2026-09-13 11:56 UTC
CompanyGramercy Extremity Orthopedics
DistributionUS

Product

GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

Reason

Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.

Identifiers

code_info
UDI-DI B471701200010, Serial Number(Expiration Date): 1022000493(10/13/27), 1022000623(10/13/27), 1022000683(10/13/27)…UDI-DI B471701200010, Serial Number(Expiration Date): 1022000493(10/13/27), 1022000623(10/13/27), 1022000683(10/13/27), 1022000243(10/13/27, 1022000313(10/13/27), 1022000913(10/13/27), 1022001213(10/13/27), 1022000403(10/13/27), 1022000383(10/13/27), 1022000883(10/13/27), 1022000353(10/13/27), 1022000663(10/13/27), 1022000733(10/13/27), 1022000793(10/13/27), 1022000983(10/13/27), 1022000993(10/13/27), 1022000373(10/13/27), 1022001193(10/13/27), 1022000523(10/13/27), 1022000283(10/13/27), 1022000223(10/13/27), 1022000363(10/13/27), 1022000413(10/13/27), 1022000453(10/13/27)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1395-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.