FDA Device recall Z-1395-2013
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 8", Part Number 2110-0525 Product Usage: Facili…
- FDA Device
- Class II
- terminated
- Published 2013-06-05
On 2013-06-05 the FDA classified a Class II recall of Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 8", Part Number 2110-0525 Product Usage: Facili… by Orthovita, citing there have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bend….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1395-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-11-10 |
| Published | 2013-06-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Orthovita |
| Distribution | US |
Product
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 8", Part Number 2110-0525 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
Reason
There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
Identifiers
| code_info | Manufacturer Part Number 2110-0525, Lot Numbers E909006R |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1395-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.