FDA Device recall Z-1394-2018
GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 Provid…
- FDA Device
- Class II
- terminated
- Published 2018-04-25
On 2018-04-25 the FDA classified a Class II recall of GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 Provid… by Ge Medical Systems Ultrasound And Primary Care Diagnostics Ll, citing under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1394-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-03-16 |
| Published | 2018-04-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ge Medical Systems Ultrasound And Primary Care Diagnostics Ll |
| Distribution | US |
Product
GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Reason
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Identifiers
| code_info | a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 |
|---|