On 2019-05-29 the FDA classified a Class II recall of Adapter for handpiece, guided, reusable, for guide sleeve 0 4.8 mm Article number: 3.04.090 - Product Usage: The adapter for handpiece, gui… by Thommen Medical Ag, citing dental adaptor does not conform to specifications. The dental coupling of the adapter can become deformed when being inserted into the implant may make it difficult to remove the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1390-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-03 |
| Published | 2019-05-29 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Thommen Medical Ag |
| Distribution | US |
Product
Adapter for handpiece, guided, reusable, for guide sleeve 0 4.8 mm Article number: 3.04.090 - Product Usage: The adapter for handpiece, guided reusable, for guide sleeve ¿ 4.8 mm is used to transfer the implant from the packaging to the patient via a plug-in connection and screwed in manually or mechanically.
Reason
Dental adaptor does not conform to specifications. The dental coupling of the adapter can become deformed when being inserted into the implant may make it difficult to remove the adapter from its counterpart (e.g. the handpiece).
Identifiers
| code_info | Lot number:17415 UDI: 07640182640705 |
|---|