RecallRadar

FDA Device recall Z-1386-2017

CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0…

On 2017-03-08 the FDA classified a Class II recall of CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0… by Maquet Datascope Cardiac Assist Division, citing maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1386-2017
ClassificationClass II
Statusterminated
Initiated2016-12-23
Published2017-03-08
Last updated2026-09-13 11:54 UTC
CompanyMaquet Datascope Cardiac Assist Division
DistributionUS

Product

CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0097 Product Usage: The CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons

Reason

Maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.

Identifiers

code_infopart number 0146-00-0097

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1386-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.