RecallRadar

FDA Device recall Z-1382-2022

Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical…

On 2022-07-20 the FDA classified a Class II recall of Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical… by Breas Medical, citing A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1382-2022
ClassificationClass II
Statusongoing
Initiated2022-06-02
Published2022-07-20
Last updated2026-09-13 11:56 UTC
CompanyBreas Medical
DistributionUS

Product

Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier

Reason

A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Identifiers

code_info
Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to…Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1382-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.