FDA Device recall Z-1382-2022
Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical…
- FDA Device
- Class II
- ongoing
- Published 2022-07-20
On 2022-07-20 the FDA classified a Class II recall of Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical… by Breas Medical, citing A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1382-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-06-02 |
| Published | 2022-07-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Breas Medical |
| Distribution | US |
Product
Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier
Reason
A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained
Identifiers
| code_info | Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to…Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1382-2022%22