RecallRadar

FDA Device recall Z-1380-2024

Baxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016

On 2024-04-03 the FDA classified a Class II recall of Baxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016 by Baxter Healthcare, citing A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1380-2024
ClassificationClass II
Statusongoing
Initiated2024-02-28
Published2024-04-03
Last updated2026-09-13 11:57 UTC
CompanyBaxter Healthcare
DistributionCT, IL, PA, TX, VA

Product

Baxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016

Reason

A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.

Identifiers

code_info
a) SPG0404, UDI/DI 00085412818672; b) SPG0406, UDI/DI 00085412818696; c) SPG0608, UDI/DI 00085412818313; d) SPG0814,…a) SPG0404, UDI/DI 00085412818672; b) SPG0406, UDI/DI 00085412818696; c) SPG0608, UDI/DI 00085412818313; d) SPG0814, UDI/DI 00085412818375; e) SPG1016, UDI/DI 00085412818429; ALL LOTS WITHIN EXPIRY

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1380-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.