FDA Device recall Z-1378-2025
Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal,…
- FDA Device
- Class II
- ongoing
- Published 2025-03-26
On 2025-03-26 the FDA classified a Class II recall of Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal,… by Oridion Medical 1987, citing neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to dis….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1378-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-02-17 |
| Published | 2025-03-26 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Oridion Medical 1987 |
| Distribution | US |
Product
Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), 006324COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology Length 7 (2.0 m), XS04620COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), XS04624COV
Reason
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Identifiers
| code_info | Part Number and Revision - REF/UDI-DI(Lot): 10129497 Rev B - 006324COV/10884521531574(D200320382, D201260758); XS04…Part Number and Revision - REF/UDI-DI(Lot): 10129497 Rev B - 006324COV/10884521531574(D200320382, D201260758); XS04620COV/10884521531529(D200322148, D200742380); XS04624COV/10884521531536(D200320376) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1378-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.