FDA Device recall Z-1378-2023
BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
- FDA Device
- Class II
- ongoing
- Published 2023-04-19
On 2023-04-19 the FDA classified a Class II recall of BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD by Bio Rad Laboratories, citing bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1378-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-02-21 |
| Published | 2023-04-19 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Bio Rad Laboratories |
| Distribution | US |
Product
BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
Reason
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.
Identifiers
| code_info | Lot # 53727; UDI GTIN: 00847865000819 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1378-2023%22