FDA Device recall Z-1378-2018
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intende…
- FDA Device
- Class II
- terminated
- Published 2018-04-25
On 2018-04-25 the FDA classified a Class II recall of MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intende… by Cook, citing the temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1378-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-01-12 |
| Published | 2018-04-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Cook |
| Distribution | US |
Product
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Reason
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Identifiers
| code_info | All Lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1378-2018%22