FDA Device recall Z-1378-2016
Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone c…
- FDA Device
- Class II
- terminated
- Published 2016-04-20
On 2016-04-20 the FDA classified a Class II recall of Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone c… by Stryker Instruments Div, citing the Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Slee….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1378-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-03-22 |
| Published | 2016-04-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Stryker Instruments Div |
| Distribution | AZ, CA, CO, CT, IL, IN, MD, MI, MN, NY, PA, SC, TN, VA, WA, WI |
Product
Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.
Reason
The Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Sleeve Assembly does not fit the Spetzler Claw" Tip, a surgical delay may occur while an alternate Tip Sleeve Assembly is retrieved.
Identifiers
| code_info | 5450-800-311 Lot Number 0545410 5450-800-313 Lot Number 0555170 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1378-2016%22