RecallRadar

FDA Device recall Z-1373-2019

CHOL+HDL+GLU test strips, REF 2412

On 2019-05-29 the FDA classified a Class II recall of CHOL+HDL+GLU test strips, REF 2412 by Polymer Technology Systems, citing some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1373-2019
ClassificationClass II
Statusterminated
Initiated2019-01-16
Published2019-05-29
Last updated2026-09-13 11:55 UTC
CompanyPolymer Technology Systems
DistributionUS

Product

CHOL+HDL+GLU test strips, REF 2412

Reason

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed

Identifiers

code_infoI808 I809 I811 I812 I814

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1373-2019%22