RecallRadar

FDA Device recall Z-1372-2019

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

On 2019-05-29 the FDA classified a Class II recall of Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412 by Polymer Technology Systems, citing some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic)….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1372-2019
ClassificationClass II
Statusterminated
Initiated2019-01-16
Published2019-05-29
Last updated2026-09-13 11:55 UTC
CompanyPolymer Technology Systems
DistributionUS

Product

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Reason

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Identifiers

code_infoQ802 Q811 Q813 Q816 Q820 Q823 Q831 Q903

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1372-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.