Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system
- FDA Device
- Class II
- ongoing
- Published 2022-07-20
On 2022-07-20 the FDA classified a Class II recall of Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system by Draeger Medical Systems, citing the firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1365-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-05-20 |
| Published | 2022-07-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Draeger Medical Systems |
| Distribution | DE, IN |
Product
Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system
Reason
The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Identifiers
| code_info | Model, country, and serial number: MS20508: El Salvador DSPN-0003 Mexico DSPM-0003, DSPH-0002 Chile DSPL-0003 S…Model, country, and serial number: MS20508: El Salvador DSPN-0003 Mexico DSPM-0003, DSPH-0002 Chile DSPL-0003 Saudi Arabia DSPJ-0003, DSPL-0004, DSPL-0005, DSPL-0006, DSPL-0007, DSPM-0001, DSPM-0002, DSPL-0008, DSPK-0002, DSPH-0003, DSPJ-0004, DSRA-0001 Kuwait DSPJ-0002, DSPK-0001 India DSPN-0001, DSPN-0002, DSRA-0002 Taiwan DSPL-0001, SPL-0002, DSPJ-0001 MS20509: Germany DSPH-0003 Qatar DSPL-0007, DSRB-0001 Spain DSPK-0002, DSPJ-0006, DSPJ-0003, DSPH-0001, DSPJ-0004, DSPK-0001, DSPL-0008 Hungary DSPF-0003 Belarus DSPJ-0005 United Kingdom DSPF-0002 Italy DSPJ-0002 |
|---|