FDA Device recall Z-1364-2022
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US…
- FDA Device
- Class II
- terminated
- Published 2022-07-20
On 2022-07-20 the FDA classified a Class II recall of AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US… by Fresenius Medical Care Holdings, citing aquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in comb….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1364-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-08-03 |
| Published | 2022-07-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Fresenius Medical Care Holdings |
| Distribution | US |
Product
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
Reason
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Identifiers
| code_info | UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1364-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.