FDA Device recall Z-1354-2018
Bard(R) Premature Infant Feeding Tube, ENFit(TM), REF EN0036400 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are…
- FDA Device
- Class II
- terminated
- Published 2018-04-18
On 2018-04-18 the FDA classified a Class II recall of Bard(R) Premature Infant Feeding Tube, ENFit(TM), REF EN0036400 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are… by C R Bard, citing BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1354-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-02-27 |
| Published | 2018-04-18 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | C R Bard |
| Distribution | US |
Product
Bard(R) Premature Infant Feeding Tube, ENFit(TM), REF EN0036400 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are intended to be used for: Decompression of stomach by suction or aspiration of gastric contents Short-term administration of term tube feeding, lavage fluid and medications Feeding Tubes with the ENFit(TM) Connector are made of a smooth plastic with a two-eyed tube used for nasogastric feeding of neonates, infants and adults as well as diagnostic and therapeutic aspiration. These products are sold single-use, sterile. The Feeding Tube with the ENFit(TM) connector provides a way to reduce the risk of enteral device misconnections. The ENFit(TM) connector: Addresses patient side connections between feeding tubes, administration sets, medication, flush and bolus feeding syringes, and other enteral devices. Does not allow connectivity with any other connector for any other clinical use.
Reason
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
Identifiers
| code_info | Lot Numbers: NGAXX216, NGAVX224, NGBN2861 |
|---|