FDA Device recall Z-1345-2016
NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This…
- FDA Device
- Class II
- terminated
- Published 2016-04-13
On 2016-04-13 the FDA classified a Class II recall of NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This… by Angiodynamics, citing procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unabl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1345-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-03 |
| Published | 2016-04-13 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Angiodynamics |
| Distribution | US |
Product
NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire
Reason
Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".
Identifiers
| code_info | Batch/Lo. 4925662 Use By 2017-02 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.