RecallRadar

FDA Device recall Z-1344-2013

Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is a rapid chromatograph…

On 2013-05-29 the FDA classified a Class III recall of Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is a rapid chromatograph… by Alere San Diego, citing alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. T….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1344-2013
ClassificationClass III
Statusterminated
Initiated2013-01-21
Published2013-05-29
Last updated2026-09-13 11:53 UTC
CompanyAlere San Diego
DistributionUS

Product

Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in the early detection of pregnancy. This test is for professional use.

Reason

Alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. The kit box incorrectly indicates the product may be used for both urine and serum sample types.

Identifiers

code_infoLot: hCG2080330

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1344-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.