FDA Device recall Z-1342-2024
MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system
- FDA Device
- Class II
- ongoing
- Published 2024-03-27
On 2024-03-27 the FDA classified a Class II recall of MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system by Angiodynamics, citing non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1342-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-01-30 |
| Published | 2024-03-27 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Angiodynamics |
| Distribution | US |
Product
MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-762
Reason
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
Identifiers
| code_info | UDI: 15051684023054 UPN: H965457621 Lot Number: Lot: 5788541 5789982 5791594 5794608 5797468 5798845 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1342-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.