FDA Device recall Z-1340-2019
Artis Zeego (material 10280959) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for si…
- FDA Device
- Class II
- terminated
- Published 2019-05-22
On 2019-05-22 the FDA classified a Class II recall of Artis Zeego (material 10280959) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for si… by Siemens Medical Solutions Usa, citing in affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1340-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-09 |
| Published | 2019-05-22 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
Artis Zeego (material 10280959) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee and Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Reason
In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.
Identifiers